Guides To Iso Cleanroom Classifications For Medical Device Manufacturing

One contamination event can cost your medical device company millions in recalls and FDA sanctions. Yet 70% of manufacturers struggle to match their products with the correct ISO cleanroom classification.

Getting your cleanroom class wrong means failed audits, production delays, and rejected shipments. We'll walk you through the exact standards you need.

ISO Cleanroom Classes Explained (ISO 5 Through ISO 8)

ISO cleanroom classifications define how many particles can exist per cubic meter of air. The lower the ISO number, the cleaner the room.

Think of it like grades. ISO 5 is your honors student with barely any particles. ISO 8 is still passing, but you'll find more contamination floating around.

Here's the breakdown:

ISO 14644-1 sets the global standard. It measures particles at two sizes: 0.5 microns and 5 microns. Your cleanroom gets classified based on these particle counts during operation.

The Four Classes You Need to Know

ISO 5 allows up to 3,520 particles (≥0.5 microns) per cubic meter. This is what you need for sterile manufacturing. Surgeries happen in ISO 5 environments.

ISO 6 permits 35,200 particles per cubic meter. You'll use this for aseptic processing areas and filling operations. It's still incredibly clean.

ISO 7 accommodates 352,000 particles per cubic meter. Most medical device assembly happens here. The gowning rooms for your ISO 5 spaces? Also ISO 7.

ISO 8 tolerates 3,520,000 particles per cubic meter. This works for non-sterile manufacturing and controlled storage areas.

ISO Class

Particles/m³ (≥0.5 μm)

Particles/m³ (≥5 μm)

Common Uses

ISO 5

3,520

29

Sterile filling, critical operations

ISO 6

35,200

293

Aseptic processing zones

ISO 7

352,000

2,930

Device assembly, gowning

ISO 8

3,520,000

29,300

Non-sterile manufacturing

The jump between classes is 10x more particles each step. That's why upgrading from ISO 7 to ISO 6 requires serious HVAC changes and tighter protocols.

iso cleanroom

Which Classification Your Device Type Requires

The FDA doesn't explicitly mandate ISO classes. But they do require appropriate environmental controls based on your device risk level.

Here's how it works: your device classification (Class I, II, or III) and intended use determine your cleanroom needs. Sterile devices need stricter controls than non-sterile ones.

Sterile Medical Devices

Class III implantable devices need ISO 5 for final assembly and sterile filling. Heart valves, pacemakers, and intraocular lenses fall here. You're putting these inside human bodies, so contamination isn't negotiable.

Class II sterile devices typically require ISO 7 for assembly with ISO 5 zones for critical operations. Surgical instruments, catheters, and wound dressings fit this category.

Sterilization method matters too. Terminally sterilized products can tolerate ISO 7 or 8 during manufacturing. Aseptically processed devices need ISO 5 from start to finish.

Non-Sterile Medical Devices

Class II non-sterile devices work fine in ISO 8. Diagnostic equipment, monitoring devices, and most external-use products don't need surgical-grade cleanliness.

Class I devices might not need classified cleanrooms at all. But many manufacturers use ISO 8 anyway to prevent quality issues and customer complaints.

Device Category

Recommended ISO Class

Examples

Class III Implants

ISO 5

Pacemakers, artificial joints

Sterile Class II

ISO 7 (ISO 5 zones)

Surgical tools, IV catheters

Aseptic Processing

ISO 5

Injectable drug delivery

Non-Sterile Class II

ISO 8

Diagnostic equipment

Class I Devices

ISO 8 or unclassified

Bandages, manual instruments

Your quality management system should document why you picked your classification. Auditors will ask.

cleanroom design

Particle Count Limits And Testing Protocols

You can't just declare your cleanroom ISO 7 and call it done. Testing proves it.

ISO 14644-2 spells out the monitoring requirements. You need baseline testing at startup, then ongoing verification to maintain classification. Skip these and your certification means nothing.

Understanding the Particle Limits

Each ISO class has maximum concentration limits at specific particle sizes. The table below shows what you're measuring:

ISO Class

0.1 μm

0.2 μm

0.3 μm

0.5 μm

1 μm

5 μm

ISO 5

100,000

23,700

10,200

3,520

832

29

ISO 6

1,000,000

237,000

102,000

35,200

8,320

293

ISO 7

-

-

-

352,000

83,200

2,930

ISO 8

-

-

-

3,520,000

832,000

29,300

Notice ISO 7 and 8 only measure larger particles. You don't need sub-micron counts for these classes.

Testing Your Cleanroom

Particle counters sample air through a laser detection system. The device pulls air through at 1 cubic foot per minute (28.3 liters).

You need multiple sampling points. The formula: square root of the cleanroom's floor area in square meters. A 100 m² room requires 10 sampling locations minimum.

Test under three conditions:

● As-built: Empty room, HVAC running

● At-rest: Equipment installed, no personnel

● Operational: Full production with workers present

Most failures happen during operational testing. People shed particles like crazy.

Guides To Iso Cleanroom Classifications For Medical Device Manufacturing 3

Design Requirements For Each Class Level

Your ISO classification dictates everything about your cleanroom design. Air handling, materials, layout, pressure differentials - they all scale with cleanliness requirements.

Getting the design wrong means you'll never hit your target particle counts. No amount of cleaning fixes bad engineering.

Air Changes and Filtration

● ISO 5 cleanrooms need 240 to 600 air changes per hour (ACH) with HEPA filters rated H14 (99.995% efficiency). You're basically creating a vertical laminar airflow that pushes particles straight down and out.

● ISO 6 requires 150 to 240 ACH. Still using H14 HEPA filters, but you can reduce the air volume slightly. The room maintains unidirectional flow in critical zones.

● ISO 7 works with 60 to 90 ACH and H13 HEPA filters (99.95% efficiency). Mixed airflow patterns work here since particle limits are more forgiving.

● ISO 8 needs only 20 to 30 ACH with H13 filters. This is where costs drop significantly. You're still filtering, just not as aggressively.

ISO Class

Air Changes/Hour

HEPA Filter Grade

Airflow Pattern

ISO 5

240-600

H14 (99.995%)

Unidirectional/laminar

ISO 6

150-240

H14 (99.995%)

Unidirectional zones

ISO 7

60-90

H13 (99.95%)

Mixed/turbulent

ISO 8

20-30

H13 (99.95%)

Mixed/turbulent

Pressure Cascades

Maintain positive pressure relative to adjacent spaces. ISO 5 rooms run at +15 Pa compared to ISO 7 corridors. This prevents dirty air from sneaking in when doors open.

Your gowning sequence should step down gradually: ISO 7 outer gown room, ISO 6 transition, then ISO 5 production area.

Surface Materials and Construction

Walls, floors, and ceilings need non-shedding materials. Epoxy-coated panels work best. Avoid drywall - it crumbles and releases particles.

Coved corners eliminate dust traps. You want seamless transitions where walls meet floors. Sharp 90-degree angles collect contamination that cleaning can't reach.

Flush-mounted lights prevent ledges. Recessed fixtures create particle collection points.

Common Compliance Mistakes To Avoid

Cleanroom compliance failures cost companies an average of $2.4 million per incident in recalls and remediation. Most of these are preventable.

The FDA catches the same mistakes repeatedly during inspections. Learn from other people's expensive lessons.

Inadequate Documentation

You tested your cleanroom once during commissioning, then never documented ongoing monitoring. Big mistake.

Auditors want continuous records. Daily particle counts, filter integrity tests, differential pressure logs, and cleaning verification. Missing documentation equals non-compliance, even if your room actually performs correctly.

Keep records for the life of the product plus three years minimum. Digital monitoring systems make this easier by auto-logging data.

Wrong Personnel Gowning Procedures

Your cleanroom design is perfect. But workers enter ISO 5 areas wearing ISO 7 gowns. Contamination central.

Each ISO class requires specific garment coverage:

● ISO 8: Lab coats, hair covers, shoe covers

● ISO 7: Full gowns, gloves, dedicated shoes

● ISO 6: Sterile gowns, double gloves, face masks

● ISO 5: Full sterile suits, goggles, no skin exposure

Train personnel on proper donning sequences. One exposed wrist negates your entire HVAC investment.

Mixing Classified and Unclassified Spaces

Opening doors directly from uncontrolled areas into ISO 7 rooms destroys your pressure cascade. You need airlocks and transition zones.

Material transfer is another weak point. Pass-through chambers with interlocked doors prevent simultaneous opening. Raw materials enter through one side, finished products exit through another.

Neglecting HVAC Maintenance

Filters don't last forever. H14 HEPA filters need replacement every 3 to 5 years, depending on usage.

Schedule preventive maintenance before filters fail. A degraded filter means your ISO 5 room is actually ISO 7. You won't know until testing reveals the problem and batches get quarantined.

Getting Your Cleanroom Classification Right

ISO cleanroom classifications aren't suggestions. They're the foundation of compliant medical device manufacturing.

Your device risk level determines your ISO class. Sterile implants need ISO 5. Non-sterile equipment works fine in ISO 8. Everything in between requires careful assessment of contamination risks and regulatory expectations.

The numbers matter:

Each class jump means 10x more particles. Design requirements scale accordingly. Air changes, filtration, pressure differentials, and personnel protocols all intensify as you move from ISO 8 to ISO 5.

Testing proves your cleanroom works. Documentation proves you maintain it. Both are non-negotiable during FDA audits.

Don't guess at your requirements. A misclassified cleanroom means failed batches, regulatory citations, and expensive retrofits. Get expert guidance upfront.

SZ Pharma specializes in compliant cleanroom design for medical device manufacturers. We'll help you match your products to the correct ISO classification and build facilities that pass qualification testing the first time.

verlieben
Worin bestehen die Unterschiede zwischen Reinräumen für die pharmazeutische Industrie und Reinräumen für die Elektronikindustrie?
keine Daten
Sich mit uns in Verbindung setzen
Kontaktieren Sie uns
Telefon: +86-18241958031/
+86-18913598640
Fax: 86-512-65488640
WhatsApp/Wechat: 008618241958031

Notiz:
Hiermit erklären wir, dass der einzige Rechtskontakt unseres Unternehmens über die oben genannte E-Mail erfolgt , Jedes andere ähnliche Suffix ist eine Betrüger-E-Mail 
Adresse: Nr. 188 Xiangwei Road, nahe der Kreuzung von Xiangwei Road und Chengyang Road (zweite Etage des Südgebäudes im Haiji Feiyue), Bezirk Xiangcheng, Stadt Suzhou, Provinz Jiangsu, China
keine Daten
Copyright © 2026 Suzhou Pharma Machinery Co., Ltd. |  Datenschutz richtlinie   Seitenverzeichnis
Kontaktiere uns
whatsapp
Wenden Sie sich an den Kundendienst
Kontaktiere uns
whatsapp
stornieren
Customer service
detect